SDM-6000 Calibration Station

SDM-6000 Calibration Station

RKI
TF: 002745-HW-RKI-045
Overview

The SDM-6000 is a versatile maintenance tool that automates the service of a GX-6000 portable gas monitor. This docking station can be used as a stand alone device for charging, calibration and bump testing, with all records stored automatically for future upload to a PC via SD card (PC software and SD card included).

DisplayTouch panel TFT liquid crystal (3.5 inch, 320×240 image element QVGA, RGB1677 million colors) Back light
LED3 color [Green, Red, Orange]. Each status is indicated by Lighting / Blinking
BUMP test

Standard setting values

AIR Flush time: 15sec, GAS time: 20sec, AIR Purge time: 5sec

Threshold value: ±50%, Auto Calibration: OFF, Auto alarm function check: ON

Auto print: OFF

Setting range

AIR Flush time: 15~120sec

GAS time: 20~120sec

AIR Purge time: 5~120sec

Threshold value: ±10~50%

Auto calibration: ON, OFF

Auto alarm function check: ON, OFF

Auto print: ON, OFF

Bump test result: Displayed on TFT liquid crystal

Calibration

Standard setting values

AIR Flush time: 30sec, GAS time: 90sec, AIR Purge time: 30sec

Auto alarm function check: ON, Auto print: OFF

Setting ranges

AIR Flush time: 15~120sec

GAS time: 20~120sec

AIR Purge time: 5~120sec

Auto alarm function check: ON, OFF

Auto print: ON, OFF

Calibration result: Displayed on touch screen display

Power (AC adaptor)100~240VAC~50/60Hz
Dimension, weight5.11”(W) x 13’(H) x 7.2”(D)in Approx. 3.7lb
Operation temperature, humidity0~104F, below 95~RH (no condensing)
USB printer portA type: Connects SDM-6000 printer (SII, DPU-S245) via USB cable
RJ-45 connectorEthernet port for a LAN connection
SD card slotCompatible with SD & SDHC cards
Joint connectorConnecting SDM-6000 (Max 10 units) and SV-3
Inlet4 inlets: GAS 1 / GAS 2 / GAS 3 / AIR
CertificationCE marking
Standard FunctionsCharging, Cal. cap fitting status monitoring, Power ON function, Alarm activation check, multi language (English, German, Japanese, Korean, Russian, Portuguese, Spanish, Italian, French), backlight adjusting, buzzer function, self-diagnosis for instruments connection
Standard accessoryOperating manual, filter, connecting plate, wall mount plate, PC controller software
SoftwarePC controller software included as standard. Windows 7 & 10 compatible
Optional accessory

SV-3 solenoid valve adds 3 additional calibration gas inputs

AC adaptor (15W output), AC adaptor (100W output)

Calibration Station utility software

Calibration gas & regulators

Tubing

Battery powered printer

SDM-6000 calibration stations are modular by design and can easily connect to each other without the need of any tubes or cables. One power source and one set of calibration equipment can provide service for up to 10 SDM-6000 calibration stations.

Up to 10 calibration stations can also be connected to one PC for automated charging, calibration, bump testing and archiving of data. Even if instruments are turned off, automated routines will power up the instruments and run as scheduled. All data collected from the PC Controlled software can also be centralized in one network location accessible by anyone with the PC Controlled software with access to the network location.