| Hardware Specifications | |
| Dimension (W x H x D) & Weight | BM3 240 x 237 x 176 / Approx 3.1 kg |
| Display / Resolution | BM3 8” / 800 X 600 |
| Parameters | BM3 ECG, Heart Rate, Respiration Rate, SpO2, Pulse Rate, Systolic BP, Diastolic BP, Mean BP, ETCO2, FiCO2, Airway Respiration Rate, 1CH-Temp |
| Trace | BM3 4 Waveforms : 2*ECG, SpO2, RR or ETCO2 |
| Indicators | Categorized Visual alarm lamp (3 priority levels) SpO2 pulse pitch tone, Battery status, External power LED |
| Interfaces | DC input connector : 18 VDC, 2.8 A LAN port for transferring data Nurse call system connection HDMI output connector USB connector |
| Battery | Rechargeable BM3 : 3250 mAh (or optional 6500 mAh) |
| Data Storage | 168 hrs (saving interval 1 min) |
| Water Resistance | IEC 60529:2013 Degress of protection provided by enclosures, IPX2 |
| ECG | |
| ECG lead & Waveforms | BM3 3-lead : 1 channel I, II, III 5-lead : 1 / 2 channels I, II, III, aVR, aVL, aVF, V |
| Heart Rate Range | Adult / Pediatric : 30 ~ 300 bpm Neonate : 30 ~ 350 bpm |
| Heart Rate Accuracy | ± 1 bpm or ± 1 %, whichever is greater |
| Sweep Speed | 6.25, 12.5, 25, 50 mm/sec |
| Filter | Diagnosis : 0.05 ~ 150 Hz Monitoring : 0.5 ~ 40 Hz Moderate : 0.5 ~ 25 Hz Maximum : 5 ~ 25 Hz |
| ST Segment Detection Range | -2.0 ~ 2.0 mV |
| Arrhythmia Analysis | BM3 : Asystole, VTach, VFib |
| Pacemaker Detection Mode | Indicator on waveform display |
| Protection | Against electrosurgical interference and defibrillation |
| Respiration Performance | |
| Method | Thoracic impedance |
| Channel Selection | BM3 : RA-LL |
| Measurement Range | 5 ~ 120 breaths per minute |
| Accuracy | ±1 breath per minute |
| Apnea Alarm | Yes |
| Temperature Performance | |
| Measurement Range | 0 ~ 50 ℃ (32 ~ 122 ℉) |
| Accuracy | 25 ~ 50 ℃ : ± 0.1 ℃ 0 ℃ to the less than 25 ℃ : ± 0.2 ℃ |
| Compatibility | 98ME04GA603 temperature probes |
| Temperature Range | Operating : 5 ~ 40 °C (41 ~ 104 °F) Storage : –20 ~ 60 °C (–4 ~ 140 °F) |
| Channels | 1CH |
| SpO2 | |
| Saturation Range | 0 ~ 100 % |
| Saturation Accuracy | 70 ~ 100 % : ± 2 digits 0 ~ 69 % : unspecified |
| Pulse Rate Range | 30 ~ 254 bpm |
| Pulse Rate Accuracy | ± 2 bpm |
| NIBP | |
| Method | Oscillometric |
| Operation Mode | Manual / Automatic / Continuous |
| Measurement Range (Adult) | Systolic : 40 ~ 260 mmHg MAP : 26 ~ 220 mmHg Diastolic : 20 ~ 200 mmHg |
| Accuracy | Mean error : less than ± 5 mmHg Standard deviation : less than 8 mmHg Meets accuracy requirements of ANSI/AAMI SP10:2002 |
| ETCO2 (Main / Sidestream) | |
| Measurement Range | 0 ~ 150 mmHg, 0 ~ 19 % |
| Accuracy | 0 ~ 40 mmHg ± 2 mmHg 41 ~ 70 mmHg ± 5 % of reading 71 ~ 100 mmHg ± 8 % of reading 101 ~ 150 mmHg ± 10 % of reading |
| Respiration Rate | 2 ~ 150 breaths per minute |
| Respiration Accuracy | ± 1 breath per minute |
The BM3 delivers compact and precise monitoring solutions with an 8-inch touch screen, 4 waveforms, and a user-friendly pop-up trend menu for low to mid acuity recovery applications.